1. Pre-Delivery & Physical Setup
Freewheel Levers
Located near the motors (Yellow or Red).
- Engage (Drive): Push levers firmly into position. The joystick will not function if disengaged.
- Disengage (Freewheel): Flip to manual push position.
Physical Checks
- Tires: Inspect for debris; verify PSI if pneumatic.
- Seating: Ensure joystick depth is adjusted to the user’s reach.
2. R-Net Controls & Driving Basics
Critical: Do not lean on the joystick during power-up. This triggers a 7-Flash Fault.
Mode Button
Toggles between Drive and Power Seating (tilt, recline, elevate).
Speed Settings
Start on the Tortoise Icon for indoor safety.
Safety Transfer
Power OFF before getting in or out to prevent accidental activation.
3. Battery Care
The "Nightly Charge" Rule
- Turn power OFF.
- Plug the 8A charger into the port beneath the joystick.
- Plug the charger into the wall outlet.
Green Light: Indicates full charge. System stops automatically to prevent overcharging.
4. Troubleshooting (Flash Codes)
| Flashes | Meaning | Action Required |
|---|---|---|
| 1 | Battery Low | Charge immediately. |
| 2-3 | Left Motor Fault | Check motor plugs. |
| 4-5 | Right Motor Fault | Check motor plugs. |
| 6 | Charger Connected | Unplug charger to drive. |
| 7 | Joystick Fault | Center joystick and restart. |
| 8 | System Fault | Power Cycle (Wait 10s). |
| 9 | Brake Fault | Engage Drive levers. |
Training Schedule: Signature Series
Setup & Programming (3.5 Hours)
Software & Hardware Integration
- Software: R-net PC Programmer & e-bx deployment.
- Hardware: Wired Dongle (D51846) for logs; Wi-Fi Dongle for monitoring.
- Status: Ensure "COMMS ACTIVE" is displayed.
Basic Programming Practices
FILE MANAGEMENT
Read from Controller FIRST to preserve factory defaults.PERFORMANCE
Tuning Tremor Damping & Low-speed Torque.Memory Seating
Joystick Method: Navigate to M1/M2 → Speed Down (3s) → Speed Up/Down (3s) to store.
Keypad Method: Hold top two left buttons (3s) until red light flashes.
Frame Adjustments
- Philosophy: Loosen bolts to slide; do not fully remove.
- Width/Depth: Use engraved guides; maintain 4” rail clearance.
- Backrest: Fixed (86°–110°) or Power Recline (up to 165°).
Diagnostics & Reset
// Common Diagnostic Logs
2C00: Battery/Voltage Fault
3B00/3C00: Motor Wiring Fault
9661-9666: e-bx Actuator Error
// E-bx Special Reset
Hold module button during power-on to reset actuator positions.
Support & Procurement
AMYLIOR TECH SUPPORT
1-888-453-0311 ext. 2
techsupport@amylior.com
DOCUMENT VERSION
Alltrack Signature v1.4